Remote Control - OPC
R&S
®
QAR
89
User Manual 1178.7200.02 ─ 04
Table 7-7: Initialization errors
Error code
Error message
Description
Solution
102
Could not load
R&S
QAR configu-
ration
Application cannot
load the configu-
ration.
Restore directory
config
in
C:\Program Files (x86)\
Rohde-Schwarz\QAROPCClientRadome
or
C:\Program Files (x86)\
Rohde-Schwarz\QAROPCClientBumper
103
Invalid operation
mode
Application config-
uration is corrupt.
Restore
C:\Program Files (x86)\
Rohde-Schwarz\QAROPCClientRadome\
config\qar_cfg.json
or
C:\Program Files (x86)\
Rohde-Schwarz\QAROPCClientBumper\
config\qar_cfg.json
104
No configuration
found for operation
mode ‘<opmode>’
No measurement
configurations
present.
Restore
C:\Program Files (x86)\
Rohde-Schwarz\QAROPCClientRadome\
QARMeasurementConfigs
or
C:\Program Files (x86)\
Rohde-Schwarz\QAROPCClientBumper\
QARMeasurementConfigs
106
Error occurred dur-
ing initialization
Replace the R&S
QAR device. Send to the ser-
vice the log files stored in
C:\Program Files (x86)\
Rohde-Schwarz\QAROPCClientRadome
or in
C:\Program Files (x86)\
Rohde-Schwarz\QAROPCClientBumper
107
R&S
QAR hard-
ware not initialized
Replace the R&S
QAR device. Send to the ser-
vice the log files stored in
C:\Program Files (x86)\
Rohde-Schwarz\QAROPCClientRadome
or in
C:\Program Files (x86)\
Rohde-Schwarz\QAROPCClientBumper
7.4.2
Measurement Errors
If errors occur during a measurement, the
Error_Flag
,
Error_Msg
and
Error_Code
are set and the device switches to
State.Results_Ready
.
Error Handling