Operating the R&S
QAR
R&S
®
QAR
65
User Manual 1178.7200.02 ─ 04
3. Start a measurement ("Measure").
4. Repeat the procedure for all three verification objects.
Verification passes if the measurement results for all three objects are within the
limits (measurement result is "OK").
5. Optional: If necessary, change the verification limits ("Options" > "Alarm Limits" >
"Difference Limit for the Verification Set").
However, we recommend to use the default limits (3
mm).
Radar Positioning Measurements