Operating the R&S
QAR
R&S
®
QAR
41
User Manual 1178.7200.02 ─ 04
Warm-up time
Let the R&S
QAR warm up for at least 90 minutes before using it.
Figure 5-4: Test setup for verification with objects
Required equipment
All required equipment is included in the delivery of the R&S
QAR-Z40.
●
Verification objects
–
1 x metal plate
–
1 x metal plate with cut-outs
–
3 x plastic plate with different thicknesses
●
Frame for verification objects
●
Allen wrench size 5
Handling the verification objects
●
Store the verification objects in the bag they were delivered in.
●
Keep the verification objects clean and without fingerprints.
●
Do not scratch the verification objects.
●
Clean the objects with a lint-free duster or a damp rag. Do not use chemicals.
Radome Measurements