Prerequisites for this command:
●
Attenuation measurement results must be available.
Return values:
<State>
1
Measured attenuation is below threshold.
0
Measured attenuation is above threshold.
Example:
//Query if attenuation is below threshold
MEAS:STAR
MEAS:OK?
Usage:
Query only
MEASurement:THReshold
<BumperType>, <Threshold>
MEASurement:THReshold?
<BumperType>
This command sets/queries the threshold value of a bumper type.
Parameters:
<Threshold>
Threshold value in dB
Parameters for setting and query:
<BumperType>
String containing the bumper type name
Example:
//Query the threshold value for the bumper type
RR_L
:MEAS:THR? 'RR_L'
//Set a new threshold value for the bumper type
RR_L
:MEAS:THR 'RR_L', 1.0
6.2.3.2
Configuration
Remote commands described elsewhere:
●
●
●
●
●
●
Remote Commands