Remote Control - SCPI
R&S
®
QAR
68
User Manual 1178.7200.02 ─ 04
Setting parameters:
<Directory>
String that contains the location of the results.
The location is always a complete directory, because measure-
ment results are made up out of several files.
Example:
//Restore measurement results
MEAS:LOAD 'c:\qar\results\measurement'
Usage:
Setting only
MEASurement:SAVE
<Description>
This command saves the current measurement results.
Prerequisites for this command
●
Measurement results must be available.
●
Define the location in which the results should be stored (
Setting parameters:
<Description>
String that is the basis of the folder name in which the R&S
QAR
saves the results.
In addition to the folder description, the R&S
QAR adds a time
stamp to the folder name. For example, a description
NewResults
would create a folder
20180101-125959_NewResults
.
Example:
//Save measurement results
MEAS:STAR
MEAS:DIR 'c:\results'
MEAS:SAVE 'NewResults'
Usage:
Setting only
MEASurement:STARt
This command initiates a measurement.
We recommend to define a timeout of at least 10 seconds to allow for the measure-
ment to finish.
Example:
//Initiate measurement
MEAS:STAR
Usage:
Event
6.2.1.2
Configuration
MEASurement:RESult:BCS:PORT
<Port>
This command defines the port the barcode scanner is connected to.
Remote Commands