2. Attach the verification plate to the frame. An arrow on the sticker indicates the cor-
rect orientation.
Figure 5-9: Placing the verification plate
Performing verification
The verification process consists of three consecutive measurements using the verifi-
cation plates.
The plates are labeled "A" to "C". Each plate has a sticker with an arrow indicating the
correct orientation and the expected attenuation value with acceptable error margin.
1. Normalize the measurement as described in
Chapter 5.2.3.1, "Normalization"
2. Put the plate labeled "A" into the support, as shown in
3. Set the "default" bumper type.
4. Initiate a measurement.
5. Check the mean attenuation against the values specified on the sticker.
6. Remove the plate from the support by pushing from the back of the plate.
Bumper Measurements