Remote Control - SCPI
R&S
®
QAR
76
User Manual 1178.7200.02 ─ 04
MEASurement:TRANsmission:NORMalization:REQuired?
This command queries the normalization state of the transmission measurement.
Return values:
<State>
0
Measurement is normalized.
1
Measurement is not normalized. In that case, you should nor-
malize the measurement before starting it.
Example:
//Query normalization state
MEAS:TRAN:NORM:REQ?
Usage:
Query only
MEASurement:TRANsmission:NORMalization:STARt
This command initiates the normalization of a transmission measurement.
We recommend to define a timeout of at least 10 seconds to allow for the measure-
ment to finish.
Example:
//Normalize transmission measurement
MEAS:TRAN:NORM:STAR
Usage:
Event
MEASurement:TRANsmission:VMOD
<State>
This command turns the verification mode for a transmission loss measurement on and
off.
Setting parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Example:
//Turn on verification mode
MEAS:TRAN:VMOD ON
6.2.2.2
Configuration
Remote commands described elsewhere:
●
●
●
●
●
●
Remote Commands