Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
217
User Manual 1178.8536.02 ─ 01
8.7.1
Configuring Standard Traces
Useful commands for trace configuration described elsewhere
●
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
●
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Remote commands exclusive to trace configuration
DISPlay[:WINDow<n>]:TRACe<t>:MODE
.........................................................................217
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:MODE:HCONtinuous
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<undef>:PRESet
.....................................219
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe]
............................................ 219
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing:APERture
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing[:STATe]
........................... 220
.......................................................................................... 220
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]
............................................................... 221
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO
..................................................... 221
........................................................................................ 222
........................................................................................... 222
...................................................................................222
.......................................................................................... 222
DISPlay[:WINDow<n>]:TRACe<t>:MODE
<Mode>
This command selects the trace mode. If necessary, the selected trace is also activa-
ted.
In case of max hold, min hold or average trace mode, you can set the number of single
measurements with
. Note that synchronization to the end of
the measurement is possible only in single sweep mode.
Suffix:
<n>
.
<t>
Parameters:
<Mode>
WRITe
Overwrite mode: the trace is overwritten by each sweep. This is
the default setting.
AVERage
The average is formed over several sweeps. The "Sweep/Aver-
age Count" determines the number of averaging procedures.
MAXHold
The maximum value is determined over several sweeps and dis-
played. The R&S
FSV/A saves the sweep result in the trace
memory only if the new value is greater than the previous one.
I/Q Analysis