Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
226
User Manual 1178.8536.02 ─ 01
CALCulate<n>:SGRam:LAYout
<State>
CALCulate<n>:SPECtrogram:LAYout
<State>
This command selects the state and size of spectrograms.
The command is available for result displays that support spectrograms.
Suffix:
<n>
.
Example:
CALC4:SPEC:LAY FULL
Shows the spectrogram in window 4. The corresponding trace
diagram is hidden.
Manual operation:
See
CALCulate<n>:SGRam[:STATe]
<State>
CALCulate<n>:SPECtrogram[:STATe]
<State>
This command turns the spectrogram on and off.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:SGR ON
Activates the Spectrogram result display.
CALCulate<n>:SGRam:THReedim[:STATe]
<State>
CALCulate<n>:SPECtrogram:THReedim[:STATe]
<State>
Activates or deactivates a 3-dimensional spectrogram for the selected result display.
Suffix:
<n>
.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
CALC:SPEC:THR:STAT ON
Manual operation:
See
I/Q Analysis