Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
197
User Manual 1178.8536.02 ─ 01
Parameters:
<NoOfSamples>
Number of samples to record.
*RST:
1001
Example:
TRAC:IQ:RLEN 256
Manual operation:
See
TRACe:IQ:SET
<FilterType>, 0, <SampleRate>, <TriggerMode>, <TriggerSlope>,
<PretriggerSamp>, <NumberSamples>
This command sets up the R&S
FSV/A for I/Q measurements.
If you do not use this command to set up I/Q measurements, the R&S
FSV/A will use
its current settings for I/Q measurements.
If the I/Q Analyzer has not been turned on previously, the command also switches to
the I/Q Analyzer.
Note:
If you use the default settings with
buffer sizes for the response data are recommended:
ASCII format: 10 kBytes
4
Binary format: 2 kBytes
Parameters:
<NORM>
<0>
Default unit: HZ
This value is always
0
.
<SampleRate>
Sample rate for the data acquisition.
Range:
100 Hz to 10
GHz, continuously adjustable
*RST:
32000000
Default unit: HZ
<TriggerMode>
Selection of the trigger source used for the measurement.
IMMediate | EXTernal | EXT2 | EXT3 | IFPower
For IMM mode, gating is automatically deactivated.
*RST:
IMM
<TriggerSlope>
Used trigger slope.
POSitive | NEGative
*RST:
POS
<PretriggerSamp>
Defines the trigger offset in terms of pretrigger samples. Nega-
tive values correspond to a trigger delay.
This value also defines the interval between the trigger signal
and the gate edge in samples.
Range:
-1399999999 to 1399999999
*RST:
0
Configuring I/Q Analyzer Measurements