Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
178
User Manual 1178.8536.02 ─ 01
OFF | 0
No data correction is performed based on the external preampli-
fier
ON | 1
Performs data corrections based on the external preamplifier
*RST:
0
Example:
INP:EGA ON
INPut<ip>:GAIN:STATe
<State>
This command turns the internal preamplifier on and off. It requires the optional pream-
plifier hardware.
For R&S
FSV/A44 or higher models, note the restrictions described in
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 15
Switches on 15 dB preamplification.
Manual operation:
See
INPut<ip>:GAIN[:VALue]
<Gain>
This command selects the gain if the preamplifier is activated (
INP:GAIN:STAT ON
,
The command requires the additional preamplifier hardware option.
For R&S
FSV/A44 or higher models, note the restrictions described in
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Gain>
15 dB | 30 dB
The availability of gain levels depends on the model of the
R&S
FSV/A.
Configuring I/Q Analyzer Measurements