Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
252
User Manual 1178.8536.02 ─ 01
Suffix:
<n>
.
<m>
Manual operation:
See
CALCulate<n>:MARKer<m>:MAXimum:NEXT
This command moves a marker to the next lower peak.
In the spectrogram, the command moves a marker horizontally to the maximum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
<m>
Manual operation:
See
CALCulate<n>:MARKer<m>:MAXimum[:PEAK]
This command moves a marker to the highest level.
In the spectrogram, the command moves a marker horizontally to the maximum level in
the currently selected frame. The vertical marker position remains the same.
If the marker is not yet active, the command first activates the marker.
Suffix:
<n>
.
<m>
Manual operation:
See
CALCulate<n>:MARKer<m>:MAXimum:RIGHt
This command moves a marker to the next lower peak.
The search includes only measurement values to the right of the current marker posi-
tion.
In the spectrogram, the command moves a marker horizontally to the maximum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
<m>
Manual operation:
See
I/Q Analysis