Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
156
User Manual 1178.8536.02 ─ 01
[SENSe:]CORRection:COLLect[:ACQuire]
<MeasType>
This command initiates a reference measurement (calibration). The reference mea-
surement is the basis for the measurement normalization. The result depends on
whether a reflection measurement or transmission measurement is performed (see
To obtain a correct reference measurement, a complete sweep with synchronization to
the end of the sweep must have been carried out. This is only possible in the single
sweep mode.
This command is only available if external generator control is active (see
SOURce<si>:EXTernal<gen>[:STATe]
Setting parameters:
<MeasType>
THRough | OPEN
THRough
"TRANsmission" mode: calibration with direct connection
between external generator and device input
"REFLection" mode: calibration with short circuit at the input
OPEN
only allowed in "REFLection" mode: calibration with open input
Example:
INIT:CONT OFF
Selects single sweep operation
CORR:METH TRAN
Selects a transmission measurement.
CORR:COLL THR;*WAI
Starts the measurement of reference data using direct connec-
tion between generator and device input and waits for the sweep
end.
Usage:
Setting only
Manual operation:
See
See
[SENSe:]CORRection:METHod
<Type>
This command selects the type of measurement to be performed with the external gen-
erator.
This command is only available if external generator control is active (see
SOURce<si>:EXTernal<gen>[:STATe]
Parameters:
<Type>
REFLection
Selects reflection measurements.
TRANsmission
Selects transmission measurements.
*RST:
TRANsmission
Configuring I/Q Analyzer Measurements