Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
174
User Manual 1178.8536.02 ─ 01
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
<ReferenceLevel>
This command defines the reference level (for all traces in all windows).
With a reference level offset
≠
0, the value range of the reference level is modified by
the offset.
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<ReferenceLevel>
The unit is variable.
Range:
see datasheet
*RST:
0 dBm
Default unit: DBM
Example:
DISP:TRAC:Y:RLEV -60dBm
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
<Offset>
This command defines a reference level offset (for all traces in all windows).
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<Offset>
Range:
-200 dB to 200 dB
*RST:
0dB
Default unit: DB
Example:
DISP:TRAC:Y:RLEV:OFFS -10dB
Manual operation:
See
" Shifting the Display ( Offset )"
8.4.2.2
Configuring the Attenuation
......................................................................................... 175
INPut<ip>:ATTenuation:AUTO:MODE
...............................................................................175
INPut<ip>:ATTenuation
<Attenuation>
This command defines the total attenuation for RF input.
If an electronic attenuator is available and active, the command defines a mechanical
attenuation (see
Configuring I/Q Analyzer Measurements