Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
154
User Manual 1178.8536.02 ─ 01
Manual operation:
See
SYSTem:COMMunicate:RDEVice:GENerator<gen>:LINK
<Type>
This command selects the link type of the external generator if the GPIB interface is
used.
The difference between the two GPIB operating modes is the execution speed. While,
during GPIB operation, each frequency to be set is transmitted to the generator sepa-
rately, a whole frequency list can be programmed in one go if the TTL interface is also
used. Frequency switching can then be performed per TTL handshake which results in
considerable speed advantages.
This command is only available if external generator control is active (see
SOURce<si>:EXTernal<gen>[:STATe]
Suffix:
<gen>
.
Parameters:
<Type>
GPIB | TTL
GPIB
GPIB connection without TTL synchronization (for all generators
of other manufacturers and some Rohde & Schwarz devices)
TTL
GPIB connection with TTL synchronization (if available; for most
Rohde&Schwarz devices)
*RST:
GPIB
Example:
SYST:COMM:RDEV:GEN:LINK TTL
Selects GPIB + TTL interface for generator operation.
Manual operation:
See
SYSTem:COMMunicate:RDEVice:GENerator<gen>:TYPE
<Type>
This command selects the type of external generator.
For a list of the available generator types see the "External Generator Control Basics"
section in the R&S
FSV/A User Manual.
Suffix:
<gen>
.
Parameters:
<Name>
<Generator name as string value>
*RST:
SMU02
Example:
//Select an external generator
SYST:COMM:RDEV:GEN:TYPE 'SMW06'
Manual operation:
See
Configuring I/Q Analyzer Measurements