Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
221
User Manual 1178.8536.02 ─ 01
Suffix:
<n>
.
1..n
Parameters:
<Mode>
VIDeo
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
*RST:
VIDeo
Example:
AVER:TYPE LIN
Switches to linear average calculation.
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]
<Detector>
Defines the trace detector to be used for trace analysis.
Suffix:
<n>
.
<t>
Example:
DET POS
Sets the detector to "positive peak".
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO
<State>
This command couples and decouples the detector to the trace mode.
Suffix:
<n>
.
<t>
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
DET:AUTO OFF
The selection of the detector is not coupled to the trace mode.
Manual operation:
See
I/Q Analysis