Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
219
User Manual 1178.8536.02 ─ 01
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<undef>:PRESet
<ResultType>
Applies predefined, commonly required trace settings to the selected window.
Suffix:
<n>
.
1..n
<w>
1..n
subwindow
<undef>
1..n
Parameters:
<ResultType>
ALL | MAM | MCM
ALL
Preset All Traces
MAM
Max | Avg | Min
MCM
Max | ClrWrite | Min
Example:
DISP:WIND3:TRAC:PRES MCM
In window 3, the traces are set to the following modes:
Trace 1: Max Hold
Trace 2: Clear Write
Trace 3: Min Hold
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe]
<State>
This command turns a trace on and off.
The measurement continues in the background.
Suffix:
<n>
.
<w>
subwindow
<t>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
DISP:TRAC3 ON
Manual operation:
See
" Trace 1 / Trace 2 / Trace 3 / Trace 4 / Trace 5 / Trace 6 "
on page 90
See
" Trace 1 / Trace 2 / Trace 3 / Trace 4 (Softkeys)"
I/Q Analysis