Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
266
User Manual 1178.8536.02 ─ 01
Return values:
<Results>
Measured voltage for I and Q component for each sample that
has been captured during the measurement.
The number of samples depends on
. In ASCII
format, the number of results is 2* the number of samples.
The data format depends on
on page 266.
Default unit: V
Example:
TRAC:IQ:STAT ON
Enables acquisition of I/Q data
TRAC:IQ:SET NORM,10MHz,32MHz,EXT,POS,0,4096
Measurement configuration:
Sample Rate = 32 MHz
Trigger Source = External
Trigger Slope = Positive
Pretrigger Samples = 0
Number of Samples = 4096
FORMat REAL,32
Selects format of response data
TRAC:IQ:DATA?
Starts measurement and reads results
TRACe:IQ:DATA:FORMat
<Format>
This command selects the order of the I/Q data.
For details see
Chapter B, "Reference: Format Description for I/Q Data Files"
Parameters:
<Format>
COMPatible | IQBLock | IQPair
COMPatible
I and Q values are separated and collected in blocks: A block
(512k) of I values is followed by a block (512k) of Q values, fol-
lowed by a block of I values, followed by a block of Q values etc.
(I,I,I,I,Q,Q,Q,Q,I,I,I,I,Q,Q,Q,Q...)
IQBLock
First all I-values are listed, then the Q-values
(I,I,I,I,I,I,...Q,Q,Q,Q,Q,Q)
IQPair
One pair of I/Q values after the other is listed
(I,Q,I,Q,I,Q...).
*RST:
IQP
TRACe:IQ:DATA:MEMory?
[<OffsetSamples>,<NoOfSamples>]
This command queries the I/Q data currently stored in the capture buffer of the
R&S
FSV/A.
Retrieving Results