Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
216
User Manual 1178.8536.02 ─ 01
Parameters:
<Time>
refer to data sheet
*RST:
depends on current settings (determined automati-
cally)
Default unit: S
Manual operation:
See
SYSTem:SEQuencer
<State>
This command turns the Sequencer on and off. The Sequencer must be active before
any other Sequencer commands (
INIT:SEQ...
) are executed, otherwise an error will
occur.
A detailed programming example is provided in the "Operating Modes" chapter in the
R&S
FSV/A User Manual.
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
The Sequencer is activated and a sequential measurement is
started immediately.
OFF | 0
The Sequencer is deactivated. Any running sequential measure-
ments are stopped. Further Sequencer commands
(
INIT:SEQ...
) are not available.
*RST:
0
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single Sequencer mode so each active measurement will
be performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
SYST:SEQ OFF
8.7
I/Q Analysis
General result analysis settings concerning the trace, markers, etc. can be configured
using the following commands. They are identical to the analysis functions in the Spec-
trum application except for the special marker functions.