Configuration
R&S
®
FSV/A3000 I/Q
Analyzer
62
User Manual 1178.8536.02 ─ 01
Calibrate Transmission
Starts a transmission type measurement to determine a reference trace. This trace is
used to calculate the difference for the normalized values.
For details, see
Chapter 4.3.4, "Calibration Mechanism"
Remote command:
Calibrate Reflection Short
Starts a short-circuit reflection type measurement to determine a reference trace for
calibration.
If both calibrations (open circuit, short circuit) are carried out, the calibration trace is
calculated by averaging the two measurements. The order of the two calibration mea-
surements is irrelevant.
Remote command:
Selects the reflection method.
[SENSe:]CORRection:COLLect[:ACQuire]
Starts the sweep for short-circuit calibration.
Calibrate Reflection Open
Starts an open-circuit reflection type measurement to determine a reference trace for
calibration.
If both reflection-type calibrations (open circuit, short circuit) are carried out, the refer-
ence trace is calculated by averaging the two measurements. The order of the two cali-
bration measurements is irrelevant.
Remote command:
Selects the reflection method.
[SENSe:]CORRection:COLLect[:ACQuire]
Starts the sweep for open-circuit calibration.
Source Calibration Normalize
Switches the normalization of measurement results on or off. This function is only
available if the memory contains a reference trace, that is, after a calibration has been
performed.
For details on normalization, see
Chapter 4.3.5, "Normalization"
Remote command:
Recall
Restores the settings that were used during source calibration. This can be useful if
instrument settings were changed after calibration (e.g. center frequency, frequency
deviation, reference level, etc.).
Remote command:
Data Input and Output Settings