Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
157
User Manual 1178.8536.02 ─ 01
Example:
CORR:METH TRAN
Sets the type of measurement to "transmission".
Manual operation:
See
See
See
[SENSe:]CORRection:RECall
Manual operation:
See
[SENSe:]CORRection[:STATe]
<State>
This command turns correction of measurement results (normalization) on and off.
The command is available after you have created a reference trace for the selected
measurement type with
[SENSe:]CORRection:COLLect[:ACQuire]
This command is only available if external generator control is active (see
SOURce<si>:EXTernal<gen>[:STATe]
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CORR ON
Activates normalization.
Manual operation:
See
" Source Calibration Normalize "
[SENSe:]CORRection:TRANsducer:GENerate
<Name>
This command uses the normalized measurement data to generate a transducer factor
with up to 1001 points. The trace data is converted to a transducer with unit dB and
stored in a file with the specified name and the suffix
.trd
under
c:\r_s\instr\trd
. The frequency points are allocated in equidistant steps between
start and stop frequency.
The generated transducer factor can be further adapted using the commands descri-
bed in the "Remote Commands > Configuring the R&S
FSV/A > Working with Trans-
ducers" section in the R&S
FSV/A User Manual.
Parameters:
<Name>
'<name>
Example:
CORR:TRAN:GEN 'SMW200A1'
Creates the transducer file
C:\r_s\instr\trd\SMW200A.trd
.
Manual operation:
See
Configuring I/Q Analyzer Measurements