Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
218
User Manual 1178.8536.02 ─ 01
MINHold
The minimum value is determined from several measurements
and displayed. The R&S
FSV/A saves the sweep result in the
trace memory only if the new value is lower than the previous
one.
VIEW
The current contents of the trace memory are frozen and dis-
played.
BLANk
Hides the selected trace.
*RST:
Trace 1: WRITe, Trace 2-6: BLANk
Example:
INIT:CONT OFF
Switching to single sweep mode.
SWE:COUN 16
Sets the number of measurements to 16.
DISP:TRAC3:MODE WRIT
Selects clear/write mode for trace 3.
INIT;*WAI
Starts the measurement and waits for the end of the measure-
ment.
Manual operation:
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:MODE:HCONtinuous
<State>
This command turns an automatic reset of a trace on and off after a parameter has
changed.
The reset works for trace modes min hold, max hold and average.
Note that the command has no effect if critical parameters like the span have been
changed to avoid invalid measurement results
Suffix:
<n>
.
<w>
subwindow
<t>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
DISP:WIND:TRAC3:MODE:HCON ON
Switches off the reset function.
Manual operation:
See
I/Q Analysis