![background image](http://html1.mh-extra.com/html/rands/fsh18/fsh18_operating-manual_774939094.webp)
R&S FSH
Saving and Recalling Settings and Test Results
1145.5973.12 2.29
E-14
Use the key to move the cursor to the right.
Insert a new letter or number at the cursor position using the alphanumeric keypad.
Press the BACK key to delete the letter or digit to the left of the cursor.
Saving Calibration Data
When performing scalar transmission or return loss measurements, the R&S FSH can store the
calibration data along with the settings and results. Saving the settings and results with calibration data
requires twice as much memory space as without it. This, of course, reduces the maximum number of
data sets that can be saved. With the R&S FSH6 and R&S FSH3 model 23 (from serial number
102314), you can now also save the calibration data for the vector measurement of the transmission or
of the reflection.
In the default state, calibration data storage is disabled.
Press the SETUP key.
Press the GENERAL softkey.
Select SAVE CAL DATA... and confirm by pressing
the ENTER key or the GENERAL softkey.
Using the rotary knob or the cursor keys, select ON
or OFF.
Confirm with ENTER.
The state for calibration data saving is entered in the SETUP menu.
When recalling data sets with stored calibration data, the R&S FSH checks whether the current
instrument temperature corresponds to the instrument temperature at the time the data was stored. If it
deviates more than 5°C, the R&S FSH displays a red dot in front of the
•
Transmission or
•
Reflection
display. Recalibration is then necessary.
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com