Saving and Recalling Settings and Test Results
R&S FSH
1145.5973.12 2.30
E-14
Recalling Measurement Results
Use the R&S FSH’s recall function to review previously saved measurement results and settings.
Press the SAVE / PRINT key.
Press the RECALL softkey.
A list of all saved data sets opens. The red selection
bar marks the last data set to be saved.
Select a data set from the list using the rotary knob.
Confirm your selection by pressing the RECALL
softkey.
The selected data set is displayed on the screen, but
the R&S FSH is not set to the settings in the data set.
You can now check the data set before its settings are
activated.
The name of the selected data set is displayed at the
lower left-hand corner of the screen.
Using the rotary knob or the cursor keys, you can scroll
through all the available data sets. The settings and
results for each data set are displayed.
You now have the following options:
Press the STATUS key to see all the instrument
settings in the selected data set. When you press
the STATUS key again, the R&S FSH returns to the
graphical display.
Press the ACTIVATE softkey to load the data set.
Press the EXIT softkey to display the list of data sets
again. Press EXIT a second time and the R&S FSH
returns to its previous settings without loading a data
set.
Press SCREEN->PRINTER to send the displayed
data set to a printer.
Pressing the ACTIVATE softkey transfers the stored trace to the R&S FSH's trace memory. The current
trace can be compared with the stored one by switching on the trace memory.
Press the TRACE key.
Press the SHOW MEMORY softkey.
The R&S FSH displays the stored trace in white and the current trace in yellow.
Note:
The trace is in the R&S FSH's trace memory. The level and frequency values are
displayed correctly only if neither the instrument's frequency setting nor its level setting
have been changed.