![background image](http://html1.mh-extra.com/html/rands/fsh18/fsh18_operating-manual_774939029.webp)
Data Sheet
R&S FSH3, R&S FSH6, R&S FSH18
1145.5973.12
TI.20
E-14
Specification
Condition
R&S FSH3
(only for model 1145.5850.23 as of serial
number 103500)
3GPP FDD code domain power BTS/Node B measurement
(only with R&S FSH-K4 1300.7633.02)
Frequency range
10 MHz to 3 GHz
Carrier frequency error
(test case 6.3 in accordance with 3GPP 25.141)
Measurement range
±
1 kHz
Measurement uncertainty
S/N > 30 dB
< 50 Hz +
S
f
ref
1)
(
T
= 20 Hz)
Total power
S/N > 30 dB
(test case 6.2.1 in accordance with 3GPP 25.141)
Measurement range
frequency > 1 MHz
20 °C to 30 °C
-60 dBm < P
total
< 20 dBm
Measurement uncertainty
-40 dBm < P
total
< 20 dBm
P
REF_LEV
-30dB < P
total
<
P
REF_LEV
+3dB
±
1.5 dB, typ. 0.5 dB
CPICH power
S/N > 30 dB
(test case 6.2.2 in accordance with 3GPP 25.141)
Measurement range
-40 dBm < P
total
< 20 dBm
P
total
-20 dB < P
CPICH
< P
total
Measurement uncertainty
- P
total
-20 dBm < P
CPICH
< P
total
±
1.5 dB, typ. 0.5 dB
P-CCPCH power
S/N > 30 dB
Measurement range
-40 dBm < P
total
< 20 dBm
P
total
-40 dB < P
PCCPCH
< P
total
Measurement uncertainty
P
total
-20 dBm < P
PCCPCH
< P
total
±
1.5 dB, typ. 0.5 dB
PSCH/SSCH power
S/N > 30 dB
Measurement range
-40 dBm < P
total
< 20 dBm
P
total
-30 dB < P
SCH
< P
total
Measurement uncertainty
P
total
-20 dBm < P
PSCH
< P
total
±
2.5 dB, typ. 1.5 dB
Symbol EVM
Measurement range
3% < EVM
symbol
< 25%
Measurement uncertainty
3% < EVM
symbol
< 10%
±
2.5% typ.
10% < EVM
symbol
< 20%
±
3.0% typ.
Residual EVM
symbol
3% typ.
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com