R&S FSH
Using the Markers
1145.5973.12 4.27
E-14
Deactivating the marker search range:
If a marker search range is activated, the LIMIT RANGE menu item in the SEARCH RANGE LIMITS
menu is highlighted in green.
Press the SET MARKER softkey to deactivate the marker search range.
Using the rotary knob or the cursor keys, select SEARCH RANGE LIMITS.
Confirm your entry with the SET MARKER softkey or the ENTER key.
Select the LIMIT RANGE menu item using the cursor keys or the rotary knob.
Deactivate the search in the limited range using the SET MARKER softkey or the ENTER key.
If you call up the SEARCH RANGE LIMITS menu again, the LIMIT RANGE menu item is no longer
highlighted.
Using more than one marker at a time (multimarker mode)
To measure different signals in a trace, the R&S FSH has the multimarker function. Up to six different
markers are available in the multimarker mode. Marker 1 measures in absolute units. Markers 2 to 6
can measure in absolute units (marker) as well as relative units (delta). The reference for delta markers
is always marker 1.
Operating sequence:
Press the MARKER key.
Press the MARKER MODE softkey.
Using the cursor keys or the rotary knob, select
MULTI MARKER... from the submenu.
Confirm your selection with the ENTER key or the
MULTI MARKER softkey.
The R&S FSH is now in multimarker mode.
Except for the MARKER DEMOD softkey that is
replaced by the MKR LIST VIEW softkey in
multimarker mode, this menu is identical to the
standard marker menu. The marker designation
contains the number of the marker in question (M
becomes M1, D becomes D2). The active marker or
delta marker is displayed with its number (e.g. M1: or
D2:) and the frequency and level at the upper-right
hand corner of the screen.