R&S ESU
Receiver Mode
4.62
Operating Manual 1302.6163.12 - 03
RUN PRE-
SCAN+FINAL
The
RUN PRE-SCAN+FINAL
softkey is starting a sequence, consisting of a pres-
can, the peak search-function and the final measurement.
Remote command:
INIT2:EMIT
RUN SCAN
4.5.5.6
Running a Scan
The
RUN SCAN
softkey starts the frequency scan with the selected settings. The
HOLD SCAN
submenu is displayed instead of the menu shown before the scan is
started.
At the beginning of the scan, R&S ESU sets up the diagram as specified in the scan
table and starts the scan in the selected mode (
SINGLE
or
CONTINUOUS
). With
SINGLE
selected, R&S ESU performs a single scan and stops at the end frequency.
With
CONTINUOUS
selected, the scan is performed continuously until it is deliber-
ately stopped.
The measurement can be interrupted with
HOLD SCAN
or stopped with
STOP
SCAN
. The two softkeys are displayed instead of the menu shown before the scan
is started.
If a transducer set is defined with points of changeover, the scan automatically stops
at the frequencies of the new subrange of the transducer set and the user may
exchange the transducer.
The following message informs the user that the limit has been reached:
TDS Range # reached, CONTINUE / BREAK
He can continue the scan at the point of change over by confirming the message
(
CONTINUE
) or he can switch off the transducer (
BREAK
).
Remote command:
INIT2
HOLD SCAN
The
HOLD SCAN
softkey interrupts the scan.
The scan stops at the frequency at which it was interrupted until it is continued with
the
CONT AT REC FREQ
or
CONTINUE AT HOLD
softkeys.
While the scan is stopped, the receiver settings can be changed, e.g. for a detailed
analysis of the recorded trace.
Remote command:
:HOLD
CONT AT REC
FREQ
With the
CONT AT REC FREQ
softkey the scan is continued at the current receiver
frequency when the receiver frequency is lower than the frequency at which the
scan was interrupted. Otherwise the scan continues at the frequency at which it was
interrupted.
The scan is always continued with the settings in the scan table.
Remote command:
:INIT2
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...