Operating Manual 1302.6163.12 - 03
4.137
R&S ESU
Instrument Functions
Analyzer Mode
Example - data section of the file
DECIM SEP
The
DECIM SEP
softkey selects the decimal separator between '.' (decimal point)
and ',' (comma) with floating-point numerals for the
function.
With the selection of the decimal separator different language versions of evaluation
programs (e.g. Microsoft Excel) can be supported.
Remote command:
FORM:DEXP:DSEP POIN
COPY TRACE
The
COPY TRACE
softkey copies the screen contents of the current trace into
another trace memory. The desired memory is selected by entering the number 1, 2
or 3.
Upon copying, the contents of the selected memory is overwritten and the new con-
tents displayed in view mode.
Remote command:
TRAC:COPY TRACE1,TRACE2
4.8.8.1
Selection of Detector
The detectors of the R&S ESU are implemented as pure digital devices. The
following detectors are available:
•
“Peak detectors (MAX PEAK and MIN PEAK)” on page 4.138
•
“Auto peak detector” on page 4.138
•
“Sample detector” on page 4.138
•
•
“Average detector” on page 4.139
•
“Quasipeak detector” on page 4.139
•
“CISPR RMS detector” on page 4.139
•
“CISPR average detector” on page 4.139
File contents
Description
Trace 1:;;
Selected trace
x-Unit;Hz;
Unit of x values:
Hz with span > 0; s with span = 0;
dBm/dB with statistics measurements
y-Unit;dBm;
Unit of y values:
dB*/V/A/
Ω
depending on the selected unit with y axis LOG or % with
y-axis LIN
Preamplifier;OFF;
Preamplifier state (OFF, if no preamplifier is available)
Transducer;OFF;
Transducer state:
OFF:No transducer active (or available)
<trd name>selected transducer
Values; 625;
Number of measurement points
10000;-10.3;-15.7
10180;-11.5;-16.9
10360;-12.0;-17.4
...;...;
Measured values:
<x value>, <y1>, <y2>
<y2> being available only with detector AUTOPEAK and containing
in this case the smallest of the two measured values for a test point.
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...