R&S ESU
Instrument Functions
R&S ESU Initial Configuration – PRESET Key
4.8
Operating Manual 1302.6163.12 - 03
Table 4-2
ANALYZER SETUP of R&S ESU
Parameter
Settings
Mode
Spectrum
Center frequency step size
0.1 ×
center frequency
RF attenuation
auto (10 dB)
Reference level
-20 dBm
Level range
100 dB log
Level unit
dBm
Sweep time
auto
Resolution bandwidth
auto (3 MHz)
Video bandwidth
auto (10 MHz)
FFT filters
off
Span / RBW
50
RBW / VBW
0,33
Sweep
cont
Trigger
free run
Trace 1
clr write
Trace 2/3
blank
Detector
auto peak
Trace math
off
Frequency offset
0 Hz
Reference level offset
0 dB
Reference level position
100 %
Grid
abs
Cal correction
on
Noise source
off
Input
RF 1, AC-coupled
Display
Full screen, active screen A
Tracking generator (only with
option R&S FSP-B9)
off
External generator 1/2 (only
with option R&S FSP-B10)
off
Preamplifier
off
Preselector
off
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...