R&S ESU
Receiver Mode
4.72
Operating Manual 1302.6163.12 - 03
The measurement detector for the individual display modes can be selected directly
by the user.
The default setting is trace 1 in the overwrite mode (
CLEAR / WRITE)
and detector
MAX PEAK
is selected, trace 2 is also in the overwrite mode (
CLEAR / WRITE)
and
detector
AVERAGE
is selected, trace 3 is switched off (
BLANK
).
The
CLEAR/WRITE, MAX HOLD, MIN HOLD, AVERAGE, VIEW
and
BLANK
soft-
keys are mutually exclusive selection keys.
SELECT
TRACE
The
SELECT TRACE
softkey activates the entry for the active trace (1, 2, 3).
Remote command:
-- (selected via numeric suffix of :
TRACe)
CLEAR/
WRITE
The
CLEAR/WRITE
softkey activates the overwrite mode for the collected mea-
sured values, i.e. the trace is overwritten by each sweep.
Each time the
CLEAR/WRITE
softkey is actuated, R&S ESU clears the selected
trace memory and starts the measurement anew.
Remote command:
DISP:WIND:TRAC:MODE WRIT
MAX HOLD
The
MAX HOLD
softkey activates the max peak detector.
R&S ESU saves the sweep result in the trace memory only if the new value is
greater than the previous one.
This is especially useful with modulated or impulsive signals. The signal spectrum is
filled up upon each scan until all signal components are detected in a kind of enve-
lope.
Pressing the
MAX HOLD
softkey again clears the trace memory and restarts the
max hold mode.
Remote command:
DISP:WIND:TRAC:MODE MAXH
VIEW
The
VIEW
softkey freezes the current contents of the trace memory and displays it.
If in the
VIEW
display mode the level display range (
GRID RANGE
) or the reference
level (
GRID MIN LEVEL
) are changed, R&S ESU automatically adapts the mea-
sured data to the changed display range. This allows an amplitude zoom to be made
after the measurement in order to show details of the trace.
Remote command:
DISP:WIND:TRAC:MODE VIEW
BLANK
The
BLANK
softkey activates the blanking of the trace on the screen.
Remote command:
DISP:WIND:TRAC OFF
SCAN
COUNT
The
SCAN COUNT
softkey activates the entry of the number of scan used in the
SINGLE SCAN mode.
The allowed range of values is 0 to 30000. The default setting is 1.
Remote command:
SWE:COUN 10
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...