TRACe Subsystem
SPURious
reads the peak list in the spurious measurement. As results a list of
frequency, level and delta to limit line values is returned. A delta limit of +200dB
indicates no limit check is active.
PHOLd reads the maximum value obtained from successive measurements.
SINGle
is possible only in form of a query for single measurements in the receiver
mode. The values of all activated detectors are transferred separated by commas
in the following order: POS, NEG, QPE, AVER, RMS. For inactive detectors, the
value and the comma are omitted at the position(s) in question.
SCAN
is only possible in form of a query during scan measurements. The number
of transmitted measurements results depend on the scan settings.
STATus
is only possible in form of a query during the scan measurements. 1 byte
status information per measurement result is transmitted:
bit 2: overrange trace1 to 3;
FINAL1
to
FINAL3
are only possible in form of a query. The final measurement
values are transmitted.
Return value:
The returned values are scaled in the current level unit. Returned FM-modulated
measurement values (activated option R&S FS-K7) are scaled in Hz.
Example:
"TRAC TRACE1,"+A$ (A$: data list in the current format)
"TRAC? TRACE1"
Characteristics:
*RST value: -
SCPI: conforming
Mode:
all
Return value:
The returned values are scaled in the current level unit. Returned FM-modulated
measurement values (activated option FS-K7) are scaled in Hz.
ASCII format (FORMat ASCII):
In ASCII format, a list of values separated by commas is returned (Comma Separated Values = CSV).
The number of measurement points is 625 (default). For more information refer to
Binary format (FORMat REAL,32):
If the transmission takes place using the binary format (REAL,32), the data are transferred in block
format (Definite Length Block Data according to IEEE 488.2). They are arranged in succeeding lists of
I and Q data of 32 Bit IEEE 754 floating point numbers. General structure of return string for a sweep
with 625 points:
#42500<meas value 1><meas value value2>...<meas value 625>
with
– #4: digits of the subsequent number of data bytes (4 in the example)
– 2500: Number of subsequent data bytes (2500 in the example)
– <meas value x>: 4 byte floating point measurement values
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...