Operating Manual 1302.6163.12 - 03
4.165
R&S ESU
Instrument Functions
Analyzer Mode
The
MKR
->
CF STEPSIZE
function is, above all, useful in the measurement of har-
monics with large dynamic range (narrow bandwidth and narrow span).
The softkey is not available in the time domain (span = 0 Hz).
Remote command:
CALC:MARK:FUNC:CST
Example
The harmonics levels of a CW carrier are to be measured at 100 MHz.
MIN
The
MIN
softkey sets the active marker/delta marker to the minimum of the selected
trace.
Remote command:
CALC:MARK:MIN
CALC:DELT:MIN
NEXT MIN
The
NEXT MIN
softkey sets the active marker/delta marker to the next higher mini-
mum of the selected trace.
Remote command:
CALC:MARK:MIN:NEXT
CALC:DELT:MIN:NEXT
NEXT MIN
RIGHT
The
NEXT MIN RIGHT
softkey sets the active marker/delta marker to the next
higher minimum right of the current marker position on the selected trace.
Remote command:
CALC:MARK:MIN:RIGH
CALC:DELT:MIN:RIGH
NEXT MIN
LEFT
The
NEXT MIN LEFT
softkey sets the active marker/delta marker to the next higher
minimum left of the current marker position on the selected trace.
Remote command:
CALC:MARK:MIN:LEFT
CALC:DELT:MIN:LEFT
[
PRESET
]
R&S ESU is set to the default setting.
[
CENTER
: 100
MHz
]
R&S ESU sets the center frequency to 100 MHz. The span
is set to 200 MHz.
[
SPAN
: 1
MHz
]
The span is set to 100 MHz.
[
MKR->
]
MARKER 1 is switched on and set to the maximum value
of the signal.
[
NEXT
]
R&S ESU switches to the submenu.
[
MKR->CF STEPSIZE
]
The step size of the center frequency setting equals the
marker frequency (100 MHz).
[
CENTER
]
The center frequency entry mode is activated.
[
Right key
]
The center frequency is set to 200 MHz. The first harmonic
of the test signal is displayed.
[
MKR->:
PEAK
]
The marker is set to the harmonic and the level of the latter
is output in the marker info field.
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...