Instrument Functions
Basic Settings
NAME
The
NAME
softkey enables the entry of characteristics in the table heading.
Name
- Enter name
A maximum of 8 characters is permitted for each name. All names must be compat-
ible with the MS DOS conventions for file names. The instrument stores all limit lines
with the
.LIM
extension.
Remote command:
CALC:LIM3:NAME "GSM1"
Domain
- Select time or frequency domain
The default setting is frequency.
Remote command:
CALC:LIM3:CONT:DOM FREQ
X-Axis
- Select interpolation
Linear or logarithmic interpolation can be carried out between the frequency refer-
ence points of the table. The ENTER key toggles between LIN and LOG selection.
Remote command:
CALC:LIM3:CONT:SPAC LIN
CALC:LIM3:UPP:SPAC LIN
CALC:LIM3:LOW:SPAC LIN
Scaling -
selection of absolute or relative scaling
The limit line can either be scaled in absolute (frequency or time) or relative units. In
receiver mode, only absolute scaling is used. Any of the unit keys may be used to
toggle between
ABSOLUTE
and
RELATIVE
, the cursor must be positioned in the X-
Scaling
or the
Y-Scaling
line.
The
RELATIVE
scaling is always suitable, if masks for bursts are to be defined in the
time domain, or if masks for modulated signals are required in the frequency
domain.
A change in domain (frequency/time) is only permitted when the data point table is
empty.
X-Scaling
ABSOLUTE
The frequencies or times are interpreted as absolute
physical units.
X-Scaling
RELATIVE
In the data point table, the frequencies are referred to the
currently set center frequency. In time domain mode, the
left boundary of the diagram constitutes the reference.
Y-Scaling ABSOLUTE
The limit values refer to absolute levels or voltages.
Y-Scaling RELATIVE
The limit values refer to the reference level (Ref Level) or,
in case a reference line is set, to the reference line.
Limit values with the unit dB are always relative values.
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...