Operating Manual 1302.6163.12 - 03
6.167
R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
6.16.5
SENSe:DETector Subsystem
The SENSe:DETector subsystem controls the acquisition of measurement data via the selection of the
detector for the corresponding trace. The measurement windows are selected by SENSe1 (screen A)
and SENSe2 (screen B).
[SENSe<1|2>:]DETector<1..3>[:FUNCtion]
APEak | NEGative | POSitive | SAMPle | RMS | AVERage
| QPEak | CAVerage | CRMS
This command sets the detector for recording measured values for the selected trace in the specified
measurement window.
• The APEak detector (AutoPeak) displays the positive and also the negative peak value of the noise
floor. If a signal is detected, only the positive peak value is displayed.
• The POSitive or NEGative detector only displays the positive or the negative peak value.
• With the Sample detector the value measured at the sampling time is displayed, whereas the RMS
value of the power measured at each measurement point is displayed with the RMS detector.
• The AVERage detector displays the power average value at each measurement point.
• The detectors QPEak for quasipeak, CAVerage for CISPR average and CRMS for CISPR-RMS
perform standard-compliant signal evaluations for the EMC test equipment.
If QPEak is selected, the video filter is automatically switched off. In addition, the couplings between
the span and RBW as well as between RBW and the sweep time are switched off. They are not
restored until another detector is selected. Accordingly, the sweep time should be set to a value that
is large enough so that the quasipeak detector can settle in completely at each measurement point.
The trace is indicated as a numerical suffix for
DETector
.
Example:
"DET POS"
'Sets the detector in screen A to "positive peak".
Characteristics:
*RST value: APEak
SCPI: conform
Mode:
A
[SENSe<1|2>:]DETector<1..3>[:FUNCtion]:AUTO
ON | OFF
This command either couples the detector in the selected measurement window to the current trace
setting or turns coupling off. The trace is selected by the numeric suffix at DETector.
Example:
"DET:AUTO OFF"
Characteristics:
*RST value: ON
SCPI: conform
Mode:
A
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...