Operating Manual 1302.6163.12 - 03
6.13
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.5.1
CALCulate:DELTamarker Subsystem
The CALCulate:DELTamarker subsystem controls the delta-marker functions in the instrument.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
In receiver mode, the markers can only be activated after a scan is performed.
CALCulate<1|2>:DELTamarker<1...4>[:STATe]
ON | OFF
This command switches on and off the delta marker if delta marker 1 is selected. If marker 2, 3 or 4 is
selected and used as a marker, it is switched to delta marker mode. If the corresponding marker is not
activated, it will be activated and positioned on the maximum of the measurement curve.
If no numeric suffix is indicated, delta marker 1 is selected automatically.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
Example:
"CALC:DELT3 ON"
'Switches marker 3 in screen A to delta marker mode.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:DELTamarker<1...4>:MODE
ABSolute | RELative
This command switches between relative and absolute frequency input of the delta marker (or time
with span = 0). It affects all delta markers independent of the measurement window.
Example:
"CALC:DELT:MODE ABS"
'Switches the frequency/time indication for all delta markers to absolute values.
"CALC:DELT:MODE REL"
'Switches the frequency/time indication for all delta markers to relative to marker 1.
Characteristics:
*RST value: REL
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:DELTamarker<1...4>:AOFF
This command switches off all active delta markers.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
Example:
"CALC:DELT:AOFF"
'Switches off all delta markers in screen A.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...