R&S ESU
IF Spectrum Analysis Mode
4.94
Operating Manual 1302.6163.12 - 03
This low average value is usually corrected in noise power measurements by a 2.5
dB factor. Therefore the R&S ESU offers the selection of linear averaging. The trace
data are delogarithmized prior to averaging, then averaged and logarithmized again
for display on the screen. The average value is always correctly displayed
irrespective of the signal characteristic.
COPY TRACE
The
COPY TRACE
softkey copies the screen contents of the current trace into
another trace memory. The desired memory is selected by entering the number 1, 2
or 3.
Upon copying, the contents of the selected memory are overwritten and the new
contents displayed in view mode.
Remote command:
TRAC:COPY TRACE1,TRACE2
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...