R&S ESU
Instrument Functions
Analyzer Mode
4.158
Operating Manual 1302.6163.12 - 03
4.8.11.5
Measurement of a Peak List
PEAK LIST
The
PEAK LIST
softkey allows the peak values of the trace to be determined and
entered in a list with 50 entries max. The order of the entries is defined by the
SORT
MODE
:
• FREQ: sorting in ascending order of frequency values (see screenshot); if span =
0, the entries are sorted in ascending order of time values
• LEVEL: sorting according to level
The search range can be restricted by means of the
LEFT LIMIT, RIGHT LIMIT
and
THRESHOLD
softkeys. The definition of the peak values can be modified using the
PEAK EXCURSION
softkey. The
MKR->TRACE
softkey in the main menu is used to
select the trace for searching peak values.
In single sweep mode, opening the list performs a single search at the end of the
sweep. The
NEW SEARCH
softkey triggers a new sweep, determines the peak val-
ues of the trace at the end of the sweep and enters them in the list.
In continuous sweep mode, the peak search on the current trace starts as soon as
the sweep starts. This allows peak list functionality on averaged or max holded
traces in continuous mode.
Use the
PEAK LIST OFF
key to delete the list from the screen.
Remote command:
INIT:CONT OFF;
CALC:MARK:TRAC 1;
CALC:MARK:FUNC:FPE:SORT X;
INIT;*WAI;
CALC:MARK:FUNC:FPE 10;
CALC:MARK:FUNC:FPE:COUN?;
CALC:MARK:FUNC:FPE:Y?;
CALC:MARK:FUNC:FPE:X?
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...