R&S ESU
Instrument Functions
Analyzer Mode
4.114
Operating Manual 1302.6163.12 - 03
RES BW AUTO
The
RES BW AUTO
softkey couples the resolution bandwidth to the selected span.
Changing the span causes automatic adjustment of the resolution bandwidth.
Automatic coupling of resolution bandwidth to span is always recommended when a
favorable setting of the resolution bandwidth in relation to the selected span is
desired for the measurement under request.
The coupling ratio is set in the
COUPLING RATIO
submenu.
The
RES BW AUTO
softkey is only available in the frequency domain (span > 0 Hz).
The softkey is deactive in the time domain.
Remote command:
BAND:AUTO ON
VIDEO BW
AUTO
The
VIDEO BW AUTO
softkey couples the video bandwidth to the resolution band-
width. If the resolution bandwidth is changed, the video bandwidth is automatically
adjusted.
The coupling of the video bandwidth is always recommended when the minimum
sweep time is required for a selected resolution bandwidth. Narrower video band-
widths require longer sweep times due to the longer settling time. Wider bandwidths
reduce the signal/noise ratio.
The coupling ratio is set in the
COUPLING RATIO
submenu.
The coupling of the video bandwidth to the resolution filter is also permitted for the
time domain display (span = 0).
Remote command:
BAND:VID:AUTO ON
SWEEPTIME
AUTO
The
SWEEPTIME AUTO
softkey couples the sweep time to the span, video band-
width (VBW) and resolution bandwidth (RBW). The sweep time is automatically
adjusted on any change in span, resolution bandwidth or video bandwidth.
The softkey is only available in the frequency domain (span >0 Hz). It is blanked in
the time domain.
The R&S ESU always selects the shortest sweep time possible without falsifying the
signal. The maximum level error compared to using a longer sweep time is < 0.1 dB.
If additional bandwidth and level errors are to be avoided, the sweep time is to be
set to three times the time offered in coupled mode.
The frequency scan rates required when using the quasipeak detector are defined in
Annex B in CISPR 16-2-x:
The
SWEEPTIME AUTO
function is designed for detectors like sample, max peak or
average. For sweeping with the quasipeak detector a manual setting of the sweep
time should be used at any time.
Remote command:
SWE:TIME:AUTO ON
Band A
Band B
Band C/D
Frequency range
< 150 kHz
150 kHz to 30 MHz
> 30 MHz
IF bandwidth
200 Hz
9 kHz
120 kHz
Frequency scan rate
20 s/kHz
200 s/MHz
20 s/MHz
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...