R&S ESU
Remote Control – Description of Commands
SOURce Subsystem
6.224
Operating Manual 1302.6163.12 - 03
Example:
"SOUR:EXT:FREQ:NUM 4"
"SOUR:EXT:FREQ:DEN 3"
'Sets a multiplication factor of 4/3, i.e. the transmit frequency of the generator is 4/
3 times the analyzer frequency.
Characteristics:
*RST value: 1
SCPI: device-specific
Mode:
R, A
SOURce<1|2>:EXTernal<1|2>:FREQuency[:FACTor]:NUMerator
<numeric_value>
This command defines the numerator of the factor with which the analyzer frequency is multiplied to
obtain the transmit frequency of the selected generator 1 or 2 in the selected window.
Example:
"SOUR:EXT:FREQ:NUM 4"
"SOUR:EXT:FREQ:DEN 3"
'Sets a multiplication factor of 4/3, i.e. the transmit frequency of the generator is 4/
3 times the analyzer frequency.
Characteristics:
*RST value: 1
SCPI: device-specific
Mode:
R, A
SOURce<1|2>:EXTernal<1|2>:FREQuency:OFFSet
<numeric_value>
This command defines the frequency offset of the selected generator 1 or 2 with reference to the
receive frequency in the selected window.
Example:
"SOUR:EXT:FREQ:OFFS 1GHZ"
'Sets a frequency offset of the generator transmit frequency compared to the
analyzer receive frequency of 1 GHz.
Characteristics:
*RST value: 0 Hz
SCPI: device-specific
Select the multiplication factor in a way that the frequency range of the generator
is not exceeded by the following formula:
if applied to the start and stop frequency of the analyzer.
This command is only valid in combination with option Ext. Generator Control
R&S FSP-B10.
Select the multiplication factor in a way that the frequency range of the generator
is not exceeded by the following formula:
if applied to the start and stop frequency of the analyzer.
This command is only valid in combination with option Ext. Generator Control
R&S FSP-B10.
F
Generator
F
Analyzer
Numerator
Denominator
-------------------------------
⋅
F
Offset
+
=
F
Generator
F
Analyzer
Numerator
Denominator
-------------------------------
⋅
F
Offset
+
=
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...