R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.18
Operating Manual 1302.6163.12 - 03
Example:
"CALC2:DELT:FUNC:FIX ON"
'Switches on the measurement with fixed reference value for all delta markers in
screen B.
"CALC2:DELT:FUNC:FIX:RPO:X 128 MHZ"
'Sets the reference frequency in screen B to 128 MHz.
"CALC2:DELT:FUNC:FIX:RPO:Y 30 DBM"
'Sets the reference level in screen B to +30 dBm
Characteristics:
*RST value: OFF
SCPI: device-specific.
Mode:
A
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
<numeric_value>
This command sets the reference point level for all delta markers in the selected measurement window
for a measurement with fixed reference point (
CALC:DELT:FUNC:FIX:STAT ON
) to the peak of the
selected trace.
For phase noise measurements (
CALCulate:DELTamarker:FUNCtion:PNOise:STATe ON
), the
command defines a new reference point level for delta marker 2 in the selected measurement window.
Example:
"CALC:DELT:FUNC:FIX:RPO:MAX"
'Sets the reference point level for the delta marker in screen A to the peak of the
trace.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:Y
<numeric_value>
This command defines a new reference point level for all delta markers in the selected measurement
window for a measurement with fixed reference point. (
CALCulate:DELTamarker:FUNCtion:
FIXed:STATe ON
).
For phase noise measurements (
CALCulate:DELTamarker:FUNCtion:PNOise:STATe ON
), the
command defines a new reference point level for delta marker 2 in the selected measurement window.
Example:
"CALC:DELT:FUNC:FIX:RPO:Y -10dBm"
'Sets the reference point level for delta markers in screen A to -10 dBm.
Characteristics:
*RST value: - (
FUNction:FIXed[:STATe]
is set to OFF)
SCPI: device-specific
Mode:
A
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:Y:OFFSet
<numeric_value>
This command defines an additional level offset for the measurement with fixed reference value
(
CALCulate:DELTamarker:FUNCtion:FIXed:STATe ON
). For this measurement, the offset is
included in the display of all delta markers of the selected measurement window.
For phase noise measurements (
CALCulate:DELTamarker:FUNCtion:PNOise:STATe ON
), the
command defines an additional level offset which is included in the display of delta marker 2 in the
selected measurement window.
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...