Operating Manual 1302.6163.12 - 03
4.207
R&S ESU
Instrument Functions
Analyzer Mode
GATE RANGES
The
GATED RANGES
softkey opens a table to configure up to 3 gate ranges for
each trace.
Remote command:
SWE:EGAT:TRACE<1..3>:COMM
SWE:EGAT:TRACE<1..3>:STAT<1..3> ON
SWE:EGAT:TRACE<1..3>:STAR<1..3> <value>
SWE:EGAT:TRACE<1..3>:STOP<1..3> <value>
SWE:EGAT:TRACE<1..3>:PER <value>
Comment
Comment string
Period
Period of the signal to be measured
Range x Start
Begin of time period to be taken into account
Range x Stop
End of time period to be taken into account
Use Range
YES / NO: Allows to temporarily disable a range
The timing values have full numerical resolution and are only rounded for display.