R&S ESU
Instrument Functions
Analyzer Mode
4.144
Operating Manual 1302.6163.12 - 03
When switched to ON (default setting), it is checked once per minute whether a fre-
quency correction for the YIG filter is required. Frequency correction is performed if
the temperature has changed by more than 3K relative to the last instance of correc-
tion.
Remote command:
CORR:YIG:TEMP:AUTO ON | OFF
CAL RESULTS
The
CAL RESULTS
softkey calls the
CALIBRATION RESULTS
table, which shows
the correction data found during calibration.
The
CALIBRATION RESULTS
table contains the following information:
– date and time of last record of correction values
– overall results of correction value record
– list of found correction values according to function/module
The results have the following meaning:
Remote command:
CAL:RES?
The
YIG CORR ON/OFF
softkey is available only on models with a frequency
range
≥
3 GHz and
≤
40 GHz.
If the instrument is operated in a temperature-controlled environment, the cyclic
frequency drift correction can – for time-critical applications – be switched off after
an operating period of
≥
30 minutes.
PASSED
calibration successful without any restrictions
CHECK
deviation of correction value larger than expected, correction
could however be performed
FAILED
deviations of correction value too large, no correction was possi-
ble. The found correction data are not valid.
ABORTED
calibration aborted
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...