R&S ESU
Instrument Functions
External Generator Control – Option R&S FSP-B10
4.342
Operating Manual 1302.6163.12 - 03
Table 4-8
Measurement accuracy levels
Accuracy
Enhancement
label
Reason/Limitation
High
NOR
No difference between reference setting and measurement
Medium
APX
(approximation)
Change of the following settings:
• coupling (RBW, VBW, SWT)
• reference level, RF attenuation
• start or stop frequency
• output level of tracking generator
• frequency offset of tracking generator
• detector (max. peak, min. peak, sample, etc.)
Change of frequency:
• max. 625 points within the set sweep limits (corresponds to
a doubling of the span)
-
Aborted
normalization
• more than 500 extrapolated points within the current sweep
limits (in case of span doubling)
At a reference level (
REF LEVEL
) of -10 dBm and at a tracking generator output
level of the same value the analyzer operates without overrange reserve, i.e. the
analyzer is in danger of being overloaded if a signal is applied whose amplitude is
higher than the reference line. In this case, either the message
OVLD
for overload
is displayed in the status line or the display range is exceeded (clipping of the trace
at the upper diagram border = Overrange).
Overloading can be avoided as follows:
• Reducing the output level of the tracking generator (
SOURCE POWER
,
NETWORK
menu)
• Increasing the reference level (
REF LEVEL
,
AMPT
menu)
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...