PAG OERLIKON AG
SERIES-300 SIP
CLOCK (350-8888)
Dismantling the housing's top
1. Remove the weighing pan [F], lift carefully the pan holder [L]
2. Dismantle the retention ring [C], take off the cover plate [A]
3. Loose the 4 screws of the cover [B], take off the cover
4. Spread the locking ring [K] and remove it
5. Take off the membrane [J]
6. Dismantle the dust-cover [G], loose the screw of the housing's top
7. Lift the housing's top [H] a bit, disconnect the flat-wire [D]
8. Take off the housing's top
Mounting the housing's top
1. Connect the flat-wire [D] to the housing's top [H]
2. Mount carefully the housing's top on the balance, fasten the screw on the housing's top
3. Pull the dust-cover [G] over the balance
4. Insert the membrane [J], mount the locking ring [K]
5. Align the membrane [J] (without any stress and waves)
6. Place carefully the cover [B] (do not move it sidewards), fasten it with 4 screws
7. Mount the cover plate [A] and the retention ring [C]
8. Bring carefully the pan holder [L] in place, mount the weighing pan [F]
A 25.02
Summary of Contents for 300 SCS series
Page 1: ...PAG OERLIKON AG The Balance of Quality SERVICE MANUAL SERIES 300 S 300 SCS 300 SIP...
Page 135: ...PAG OERLIKON AG SERIES 300 S SCS SIP MAIN BOARD 300 7218 010 HAUPTPRINT 300 7218 010 C 01 03...
Page 150: ...PAG OERLIKON AG SERIES 300 S SCS SIP MAIN BOARD 300 7235 010 HAUPTPRINT 300 7235 010 C 02 06...
Page 159: ...PAG OERLIKON AG SERIES 300 S SIP ANALOGUE BOARD 300 7219 010 ANALOGPRINT 300 7219 010 C 03 01...
Page 162: ...PAG OERLIKON AG SERIES 300 SCS ANALOGUE BOARD 300 7238 010 ANALOGPRINT 300 7238 010 C 04 01...
Page 194: ...SERIES 300 S SCS SIP PAG OERLIKON AG CALIBRATION INSTRUCTIONS D 04 11 Final control...