PAG OERLIKON AG
SERIES-300 S / SCS / SIP
MAIN BOARD 300-7218-010
HAUPTPRINT 300-7218-010
External Switch
S3
EPROM
RAM
Calibration locked
External Tare
External On / Off
External Mode
Main Board Label
EPROM Label
On
Off
Standard Settings
SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
Switch 1 On
-->
IEC625 / IEEE488
Off
-->
disactivated
Switch 2 On
-->
USA date
Off
-->
Europe date
Switch 3 On
-->
Zero-check (shows ERROR 8)
Off
-->
disactivated
Switch 4 On
-->
shows
not
'--OFF-- Power' during start-up
Off
-->
shows '--OFF-- Power' during start-up
Switch 5 On
-->
Real Time Clock
Off
-->
disactivated
Switch 6 On
-->
with Internal Calibration
Off
-->
disactivated
Switch 7 On
-->
-
Off
-->
-
Switch 8 On
-->
Weighing Mode
Off
-->
Service Mode
DIL-Switch Setting
Software no.
series 300 S = K3x
series 300 SCS = K3x
K3x
"Software no."
EPROM-Labeling
For the function of a PRECISA balance it is absolutely essential that the software (EPROM) is
corresponding with the main board. For this reason the code on the label of the EPROM (the first three
figures) has to be the same as the one marked on the label of the main board.
Example:
"Board no."
K3x
C 01.06
Summary of Contents for 300 SCS series
Page 1: ...PAG OERLIKON AG The Balance of Quality SERVICE MANUAL SERIES 300 S 300 SCS 300 SIP...
Page 135: ...PAG OERLIKON AG SERIES 300 S SCS SIP MAIN BOARD 300 7218 010 HAUPTPRINT 300 7218 010 C 01 03...
Page 150: ...PAG OERLIKON AG SERIES 300 S SCS SIP MAIN BOARD 300 7235 010 HAUPTPRINT 300 7235 010 C 02 06...
Page 159: ...PAG OERLIKON AG SERIES 300 S SIP ANALOGUE BOARD 300 7219 010 ANALOGPRINT 300 7219 010 C 03 01...
Page 162: ...PAG OERLIKON AG SERIES 300 SCS ANALOGUE BOARD 300 7238 010 ANALOGPRINT 300 7238 010 C 04 01...
Page 194: ...SERIES 300 S SCS SIP PAG OERLIKON AG CALIBRATION INSTRUCTIONS D 04 11 Final control...