SERIES-300 S / SCS / SIP
PAG OERLIKON AG
ERROR MESSAGES
D 05.01
Error messages
1. Operating error
1.1 Error disappears with correct use
- ERROR 1
- no programme cover on the multifunction box
- ERROR 2
- no diplay possible in this unit (GN, ct, C.M.)
- reference value too small or negative (percentage programme)
- test value deviation g/- 50% from reference value (statistics programme)
- reference value s 100 digits (statistics programme)
- density of the solid-state body greater 25 g/ccm (density programme)
- at least one test value in the overload or underload area (animal weighing
programme)
- ERROR 3
- reference value too small or negative (counting programme)
- less than two test values stored (statistics programme)
- density of the liquid smaller 0.5 g/ccm or greater 2 g/ccm (density programme)
1.2 Error correctable by corresponding measures
- ERROR 7
- non-permissible extension connected (disconnect from MF-box bush)
- ERROR 8
- zero point outside tolerance limits (empty balance pan)
- ERROR 9
- calibration constant smaller 0.99 or greater 1.01, balance stops (disconnect
balance, repeat calibration operation)
2. Error in normal operation
2.1 Fatal errors
(are displayed on power-up -> quit with tare key)
- ERROR 10
- linearisation factors destroyed (re-linearise balance)
- ERROR 11
- calibration constant destroyed (re-calibrate balance)
2.2 Fatal errors
(are displayed on power-up -> balance stops)
- ERROR 13
- balance model code destroyed during operation -> disconnect and reconnect
balance)
- ERROR 14
- invalid balance model code (enter new model code)
- ERROR 16
- internal RAM defective (replace microprocessor)
- ERROR 17
- external RAM defective (replace RAM)
- ERROR 18
- EPROM defective (replace EPROM)
- ERROR 19
- wrong EPROM (install correct EPROM)
2.3 Diagnostic error for service
(tolerable errors, are displayed in test programme 1
-> quit with tare key)
- ERROR 21
- temperature factors destroyed (fine temperature compensation inactive)
3. Errors in the adjustment test
3.1 Errors in the temperature compensation (correctable by new temperature compensator)
- ERROR 50
- temperature difference smaller 100 digits or temperatures stored in wrong order
- ERROR 51
- limiting value for zero point or scale correction exceeded
- ERROR 52
- not all test values necessary for calculating temperature curve available
- ERROR 53
- increase in the zero point or scale too great
3.2 Initial loading error
(correctable by changing load)
- ERROR 60
- minimum transformer area fallen below
- ERROR 61
- maximum transformer area exceeded
Summary of Contents for 300 SCS series
Page 1: ...PAG OERLIKON AG The Balance of Quality SERVICE MANUAL SERIES 300 S 300 SCS 300 SIP...
Page 135: ...PAG OERLIKON AG SERIES 300 S SCS SIP MAIN BOARD 300 7218 010 HAUPTPRINT 300 7218 010 C 01 03...
Page 150: ...PAG OERLIKON AG SERIES 300 S SCS SIP MAIN BOARD 300 7235 010 HAUPTPRINT 300 7235 010 C 02 06...
Page 159: ...PAG OERLIKON AG SERIES 300 S SIP ANALOGUE BOARD 300 7219 010 ANALOGPRINT 300 7219 010 C 03 01...
Page 162: ...PAG OERLIKON AG SERIES 300 SCS ANALOGUE BOARD 300 7238 010 ANALOGPRINT 300 7238 010 C 04 01...
Page 194: ...SERIES 300 S SCS SIP PAG OERLIKON AG CALIBRATION INSTRUCTIONS D 04 11 Final control...