SERIES-300 S / SCS / SIP
PAG OERLIKON AG
CALIBRATION INSTRUCTIONS
Calibration instructions
D 04.01
3
2
4
22
1
1 2 3 4 5 6 7
8
1. Put the switch 8
(1)
in OFF position
(service mode)
2. Level the balance with the levelling screws
(13/B02.01)
, connect balance to the power supply
3. Adjust the balancing
- Connect the connecting cable [14] to the plug (2), connect the voltmeter to the connecting cable [14]
- The capacity readings of the voltmeter must show in the end-positions of the balance arm (12/B04.01)
about the same voltage (one time positive, one time negative) in the end positions of the scale balance
(12/B04.01)
- If the difference is too great, the position of the balance arm can be changed with the stopper (22/B04.01)
- The value of the balancing voltage is finally set with the trimmer (3) to +/- 1.5V
- Remove the connecting cable [14] and the voltmeter
4. Initial loading control
- The initial loading Error 60 (falling below minimum transformer area) with the weighing pan put on without
load and Error 61 (maximum transformer area exceeded) with the weighing pan put on with full load
should not appear
- The correction of the initial loading with jumpers (4) as per separate list {request specific type list from
Customer Service Dept.}
Summary of Contents for 300 SCS series
Page 1: ...PAG OERLIKON AG The Balance of Quality SERVICE MANUAL SERIES 300 S 300 SCS 300 SIP...
Page 135: ...PAG OERLIKON AG SERIES 300 S SCS SIP MAIN BOARD 300 7218 010 HAUPTPRINT 300 7218 010 C 01 03...
Page 150: ...PAG OERLIKON AG SERIES 300 S SCS SIP MAIN BOARD 300 7235 010 HAUPTPRINT 300 7235 010 C 02 06...
Page 159: ...PAG OERLIKON AG SERIES 300 S SIP ANALOGUE BOARD 300 7219 010 ANALOGPRINT 300 7219 010 C 03 01...
Page 162: ...PAG OERLIKON AG SERIES 300 SCS ANALOGUE BOARD 300 7238 010 ANALOGPRINT 300 7238 010 C 04 01...
Page 194: ...SERIES 300 S SCS SIP PAG OERLIKON AG CALIBRATION INSTRUCTIONS D 04 11 Final control...