23. Assembly of the safety device
(23/B04.01)
- Remove 2 auxiliary pins D4 [2] from the weighing cell
- Undo 2 screws [10], remove the fixing block [4] to one side
- Screw the special screw (17/B04.01) carefully into the balance arm
- Fasten the safety device with the screw (29/B04.01), the washer (5/B04.01) and the spring washer
(6/B04.01) on the chassis (19/B04.01) {do not tighten}
- Turn the centring bush [8] upside down above the special screw (17/B04.01) {the safety device is
centred}
- Tighten the screw (29/B04.01), remove the centring bush [8]
- Same assembly procedure for the second safety device
- Connect the flat wire of the sensor mechanism to the main board
24. Further procedure is described on pages D 04.01-05.01