27
DMC Repair Manual
Parts Replacement
Anti-Alias Optical Filter Assembly (Figure 2-7)
Removal
1. Remove the bottom extrusion.
2. Remove the front panel and optics assembly.
3. Place the front panel and the optics assembly onto an anti-static mat.
4. Remove the three (3) screws securing the optics assembly to its front panel supports.
5. Remove the CCD retainer and the backing pad.
6. Lift off the CCD imager from the optical mount being careful not to scratch the CCD imager.
7. Remove the optical mount and shutter asssembly from the front panel and place it into the
optical mount fixture.
Note: Make sure that the optical mount is inserted into the optical mount fixture with the
thread side facing up and the motor on the shutter facing down.
Figure 2-7. Removing anti-alias optical filter assembly
Optical Filter Assembly
Optics
Assembly
Optical Mount
Spanner Nut
Anti-Alias
OpticalFilter