57
DMC Repair Manual
Testing & Adjustment
7. Set calibration test parameters as follows: (Test parameters located at left side of
displayed calibration test panel.)
•
Do no tests to OFF
•
Do all tests to ON
•
Load defaults to OFF
Note:
Usually Do all test to ON is selected. If only a specific test/s needs to be run,
set Do all test to OFF and then click on the specific test/s.
- indicates that the particular test is not selected.
- indicates that the particular test is selected.
8. Click on the white arrow. Arrow changes to black indicating that the test program is
running.
9. Click on the Start button to automatically start the selected LabView VI test
program. A message appears prompting you to modify the configuration tag.
10. Click on the NO button. A message appears prompting you to modify the storage code
tag .
11. Click on the NO button. The EEPROM Test Tool.vi window appears.
12. Set Panel Defaults button to Full Test.
Note:
If only a specific test/s was selected in step 7 of this test procedure, click on
Cancel button for partial test.
Once panel defaults are selected, a message appears prompting you to set the light box
to 50 Cd/Ft
2
(candles per foot).